A modified proportional source method used for measuring the dead time of a radiation detecting system
|更新时间:2022-08-16
|
A modified proportional source method used for measuring the dead time of a radiation detecting system
Nuclear TechniquesVol. 09, Issue 1, (1986)
作者机构:
中国原子能科学研究院
作者简介:
基金信息:
DOI:
CLC:
Received:19 January 1984,
Published:1986-01
稿件说明:
移动端阅览
Changmao Chen, Jianlun Xie. A modified proportional source method used for measuring the dead time of a radiation detecting system[J]. Nuclear techniques, 1986, 09(1): 50-52,63.
DOI:
Changmao Chen, Jianlun Xie. A modified proportional source method used for measuring the dead time of a radiation detecting system[J]. Nuclear techniques, 1986, 09(1): 50-52,63.DOI:
A modified proportional source method used for measuring the dead time of a radiation detecting system
This paper describes a method of measuring the dead time of a radiation detecting system using two radiation sources with the same spectrum and structure but different intensity. It can be used to precisely determine the dead time independent from count rates
and the approximate values of the relative intensities of the sources can be obtained.