The measurement of Cerenkov radiation of 32P in ethanol system and an improvement on raising efficiency
|更新时间:2022-08-16
|
The measurement of Cerenkov radiation of 32P in ethanol system and an improvement on raising efficiency
Nuclear TechniquesVol. 08, Issue 1, (1985)
作者机构:
上海第一医学院
作者简介:
基金信息:
DOI:
CLC:
Received:19 December 1983,
Published:1985
稿件说明:
移动端阅览
Hong Zhang, Helin Dong, Baofen Zheng. The measurement of Cerenkov radiation of 32P in ethanol system and an improvement on raising efficiency[J]. Nuclear techniques, 1985, 08(1): 44-46,75.
DOI:
Hong Zhang, Helin Dong, Baofen Zheng. The measurement of Cerenkov radiation of 32P in ethanol system and an improvement on raising efficiency[J]. Nuclear techniques, 1985, 08(1): 44-46,75.DOI:
The measurement of Cerenkov radiation of 32P in ethanol system and an improvement on raising efficiency
The common methods and conditions of measuring Cerenkov radiation of
32
P by liquid scintillation counters have been reported. The efficiency is higher in ethanol system than in water. By adding a little scintillation solution to ethanol system
the efficiency can be further raised; meantime the system remains basically unquenched by chemical substance. lf the wave length shifter is put into ethanol system instead of scintillation solution
the chemical substance can result in an apparent quenching. The quenching effect of colour in ethanol system containing scintillator is shown.