This paper describes a measurement device for determination of the uranium content by delayed neutron counting. The advantages of this device are as follows: (1) high sensitivity
the detections limit for natural uranium and
235
U is 0.05 ppm and 0.4 ppb respectively
(2) no blank error
γ
balance problem and other interferences. (3) high accuracy
for 1 ppm natural uranium
the error of the determination is less than 4.3%. (4) simplicity
the analyzing time for one measurement is three minutes. (5) It can automatically analyse large amount of samples and immediately give its results. (6) samples are nondestructive. It can determine ten to twenty thousand samples per year. This equipment is suitable for determination of large amount of geological explorative samples.
关键词
Keywords
references
S. Amiel et al ., LA -621 , 1961 .
F. F. Dyer , USAEC Report ORNL -3342 , 1962 .
Report DPST -75-133-1, -2, -3, -4 , 1975 .
S. E. Biney et al ., Nucl. Instr. Meth ., 154 , 413 ( 1978 ).
N. N. Papado Poulos , J . Radiochem ., 61 , ( 1-2 ), 81 ( 1981 ).
H. T. Millard et al ., ibid , 72 ,( 1-2 ), 489 ( 1982 ).
B. J. Rosenberg , ibid , 62 , ( 1-2 ), 145 ( 1981 ).
P. Schhechte , ibid , 62 , 1-2 , 55 ( 1981 ).
P. C. Ernst et al ., ibid. 70 , ( 1-2 ), 527 ( 1982 ).