shenggang Li, Guanjun Yuan, Peikun Weng. A Ge(Li)-Ge(Li) list mode multiparameter system for E-E-T coincidence[J]. Nuclear techniques, 1983, 06(4): 15-18,73,74.
本文介绍研制的计算机控制的列表方式参数Ge(Li)-Ge(Li)符合系统的实验装置、定时特性、数据处理以及应用举例,关于列表方式数据获取与分析方法的工作特点,这里不再赘述。
For many years
the coicidence measurement techniques have often played an important role in the experiments of nuclear physics. Since the invention of the Ge(Li) detector
wihich has an excellent energy resolution and larger volume
the coincidence methods have greatly been improved. As the Ge(Li) detector has larger capacitance and small pulse-height and various rise time of pulse
its timing seems more difficult.
A Ge(Li)-Ge(Li) list mode multiparameter system with the Scorpio-3000 multichannel computer system and two large Ge(Li) detectors (one is 130 cm
3
the other 100 cm
3
) has been set up and adjusted.
The system is capable of processing 1000 events/s and E-E-T analysis in 4096 × 4096 × 1024 channels. The play back sorting of a 2500 feet magnetic tape is completed in 90 minutes
and the work can be done with the same computer on the spot
The time resolution is 30 ns at 1.33 MeV γ ray and true/false coincidence ratio is 65:1.
N. S. Kendrick Jr and D. A. Mcclurc , Nucl. Instr. Meth ., 121 , 573 ( 1974 ).
W. D. Ruhter , UCRL -52360 ( 1977 ).
D. N. Pocnaru et al . , Trans. Nucl. Sci ., NS-17 , 3 , 176 ( 1970 ).
Bernward W. Huber , Nucl. Instr. Meth ., 114 , 509 ( 1974 ).
Traverse et P. Berthe , Nucl. Instr. Meth ., 148 , 317 ( 1978 ).