众所周知,1950年勒文(H. D. Levine)等用计数管做成轻便剂量率仪,计数管便在剂量测定应用中得到发展。在直流供电下,特别是在高照射率时,由于计数管死时间的影响,其计数率
N
与照射率
P
之间不是线性关系。例如J405型的线性测量上限只在100~200R/h左右,测到1000R/h的非线性误差可达45%左右;又如J404型的线性上限只在1~2R/h左右,测到10R/h的非线性误差达35%左右。
Described here is a simple circuit used as counter pulsation power supply with good stability and reliability. It has good linearity of dose character