符合吸收曲线效率外推法,测定短寿命的医用同位素
131
I、
198
Au绝对强度。测量结果分别对本底、死时间、分辨时间进行了修正,同时还考虑了短半衰期引起的修正。对
131
I和
198
Au的最大测量误差分别为0.57%和1.7%。本装置稍作改变后,还测量了具有电子俘获衰变的核素。
The absolute intensity of medical isotopes
131
I and
198
Au with a short half-life has been measured by the efficiency extrapolation method of 4
πβ-γ
coincidence absorption curve. The result of the measurement has been corrected for background
dead time
resolving time respectively. The correction for the short half-life has also been considered. The maximum error of the measurement are 0.57% and 1.7% for
131
I and
198
Au respectively. After a little change of the apparatus
we have measured the nuclide with electron capture decay.