浏览全部资源
扫码关注微信
采用CaSO
4
(Tm)为热释光磷光体,葡萄糖为反冲质子辐射体,用CaSO
4
(Tm)和CaSO
4
(Tm)—葡萄糖剂量计对测量T(
D
n
)
4
He反应产生的14MeV快中子吸收剂量,在0.1~500rad进行了试验,得到该剂量计对的快中子灵敏度为9×10-2[rad/每rad快中子]的结果。方法的最大误差为15%,初步进行了该剂量计对的方向性试验,在一定误差范围内可以认为方向性的影响不是严重的。
The absorbed dose of 14MeV fast neutron generated by T (D
n)
4
He reaction was measured by the CaSO
4
(Tm) and CaSO
4
(Tm) -glucose dosimeter pair with CaSO
4
(Tm) as thermoluminescence phosphor and with glucose as emitter of recoiled proton. The measurements made at 0.1 ~500 rad had a sensitivity of 9×10
-2
[rad/rad fast neutron] using dosimeter pair mentioned above. The maximum error of this method was 15%. A preliminary investigation for effect of direction of the dosimeter pair showed that it was not serious in a given error range.
E. Blum et al . , Health phys ., 30 , 257 ( 1976 ).
E. Blum et al . , Dosimetry in Agriculture, Industry, Biology and Medicine ., 1973 , P. 277 .
NCRP 第25号出版物
冯嘉林 等 , 核技术 , 4 , 81 ( 1979 ).