本文介绍的大面积低本底α闪烁计数器,其面积为111cm
2
,4
x
效率为93.8%,平均本底为0.066cpm。取
u
α
=
u
β
=2,即取两者的置信限为97.73%,测量100分钟,其最小可探测真实活性为9.58×10
-14
Ci;最小可探测真实比活性在0.2g灰样时为5.8×
M
×10
-13
Ci/kg。同时简要地说明了仪器的设计和使用原则。
A large area low background
α
scintillation counter is described. Its effective detecting area is 111cm
2
and counting efficiency with
4π
geometry is 93.8%. The average background count rate is 0.067cpm. If
U
α
=U
β
=2 (i. e. both confidence levels are 97.73%) and measuring time is 100 minutes
the minimum detectable real activity is 9.58×10
-14
Ci and in an ash specimen of 0.2g the minimum detectable real specific activity is 5.8×10
-13
Ci/kg.
The design principle and operation of this instrument are given in brief.
H. L. Butler and W. F. Splichal , Health Phys. , 12 , 1627 ( 1966 ).
W. Mcklveen John , Radiat. Res. , 66 , 199 ( 1976 ).
W. J. Mcdowell and J. F. Weiss , Health Phys. , 32 , 73 ( 1977 ).
N. I. Sax and D. Rosi , Anal. Chem. , 32 , 1900 ( 1960 ).
D. E. Watt and D. Ramsden , High Sensitivity Counting Techniques , Pergamon Press Oxford. London 1964 , P. 1 .
R. Boulenger and E. Gourski , Proc. 2nd Int. Conf. on peaceful Uses of Atomic Energy , Geneve , Vol.23 , 1958 , P. 362 .
赖祖武 等 , 核物理电子学方法 , 上海科技出版社 , 1961 年, 第116页 .
B. De' Flagge and O. R. Harris , Rev, Scient, Instr. , 26 , 619 ( 1955 ).
苏州医学院 , 应用数理统计方法 , 1976 年.