用台式同位素源X射线荧光分析仪测定水泥样品中的CaO和Fe
2
O
3
含量,分析精确度分别为0.14%和0.02%。仪器分析值与化学分析值对照,两者偏差在±0.25%CaO内的约占80%,在±0.10%Fe
2
O
3
内的占90%以上。用仪器测定 "出磨水泥"及入磨原料中的CaO以后,便可计算出其中的熟料掺入量,偏差小于1.5%。
A desk instrument of radioisotope X-ray fluorescence analysis is described
which has been used to determine the content of CaO and Fe
2
O
3
in cement specimens
with an precision of 0.14% and 0.02% respectively. The comparison between the results of instrumental and chemical analysis shows that
the specimens of which both results are agreeable within ±0.25% deviation for Cao and within ±0.10% deviation for Fe
2
O
3
amount to about 80% and above 90% of the total respectively. After the determination of CaO content in the "ground cement" and its raw material
the examination of clinker component can be performed
with an error less than 1.5%.
徐君权 等 , 核技术 , 1 , 57 ( 1978 ).
徐君权 等 , 用于X射线荧光分析的密封式正比计数管,中国科学院核探测器和核电子学交流会会议资料 , 1979 年.
C.H. Anderson et al ., Advances in X-ray Analysis , Vol.17 , Plenum Press , 1973 , P. 214 .
J. A. Cooper et al ., Advances in X-ray Analysis , Vol.19 , Plenum Press , 1976 , P. 213 .